Slit
Lamp Illumination Types
Associated Ocular Conditions And
Slit Lamp Examination Procedures
Direct
Illuminations:
1.) Diffuse:
Diffuse illumination or "wide beam" illumination deserves
a short separate discussion from the other types of
illuminations. The term diffuse has been carried over
from earlier writings when slit lamps had either
diffusing filters or independent racking microscopes.
This allowed each to be independently focus on different
structures. Most of today's slit lamp biomicroscopes have
their light sources and microscope coincident to one
another and are focused on the same structure at the same
time. Diffusing filters are still found in some slit
lamps and are used in photographing the anterior segment
of the eye. "Wide beam" illumination is the only type
that has the light source set wide open. Its main purpose
is to illuminate as much of the eye and its adnexa at
once for general observation.
A wide,
un-narrowed, beam of light is directed at the cornea from
an angle of approximately 45 degrees. Position the
microscope directly in front of the patient's eye and
focus on the anterior of the cornea. Low to medium
magnification
(7 - 16x) should be used which allows the observer to
view as many of the structures as possible. When viewing
the eye with achromatic light one should note, on gross
inspection, any corneal scars, irregularities of the
lids, tear debris, irregularities of Descemet's membrane
or pigmentary changes found in the epithelial layer, etc.
These findings are investigated more thoroughly with
other types of illumination.
With the
aid of the cobalt blue filter and either fluorescein
sodium or Fluoresoft® permit the evaluation of
bearing, movement, positioning of contact lenses. Using
the cobalt blue filter and fluorescein sodium this is a
reasonable illumination for assessing a patient's tear
break up time (TBUT), dark drying areas of the
epithelium. Staining of the cornea and conjunctiva can
also be assessed. Fluorescein sodium dye will stain the
cornea and conjunctiva any time the epithelium is
compromised. Fluorescein dye does not stain epithelial
cells themselves, but pools within the intercellular
defects thus highlighting the damaged area. Fluorescein
staining is relatively nonspecific, occurring with any
condition affecting epithelial integrity. Do not confuse
"negative staining" for true TBUT. Negative staining is
often seen when checking TBUT or evaluating an area that
has stained in the past. "Negative staining" results
because of irregularities in the corneal epithelium.
These dark areas are where the tears separate quickly
rather than stain tissue. Examples: Map and dot
dystrophies, micro-cystic edema, healing but still rough
and not smoothly healed abrasions, etc. The cobalt blue
filter is also helpful in detecting Fleischer's Ring or
Line in keratoconus and Hudson Stahli's Line in older
patients.
Using Rose
Bengal and no filters will show a pink staining of
epithelial cell damaged tissue as in cases of
keratoconjunctivitis Sicca, herpatic lesions of the lids
and cornea and other ulcer margins. Other types of
illumination are better for evaluating the degree and
depth of any staining. See color plates in Dr. Casser's
book.
Diffuse,
wide-beam, illumination together with the red free
(green) filter is helpful when viewing the bulbar
conjunctiva, and episcleral blood vessels. With the aid
of the red free filter small hemorrhages, aneurysms and
engorged vessels stand out. It is not difficult to
differentiate between conjunctival, episcleral, and
scleral injection. Conjunctival vessels are obviously
fairly superficial and are movable upon friction from the
eyelid, whereas less superficial and deep vessels show
minimal to no movement with the overlying conjunctiva.
Deeper episcleral injection may also appear somewhat
darker and give an overall purplish hue. (Abelson et al)
proposed a standardized grading system for judging the
different types of injection.

A) Ciliary Injection B) Episcleral Injection C)
Conjunctival Injection
|
Grade
|
Severity
|
|
0
|
White and Quiet
|
|
1/2
|
Slight, Usually Normal
|
|
1 to 1+
|
Mild
|
|
2 to 2+
|
Moderate
|
|
3 to 3+
|
Severe
|
Adapted and modified from Abelson, M. et
al.
|
|
2.)Direct
Focal
a.) Optic Section: Optic section is used
primarily to determining the depth or elevation
of a defect of the cornea, conjunctiva or
locating the depth of an opacity within the lens
of the eye.
With
the optic section as mentioned above, it is
possible to detect changes in corneal and
conjunctival thicknesses, to assess depths of
foreign bodies, scars and opacities, to estimate
the anterior chamber depth and to identify the
anatomical location of cataracts within the
crystalline lens. The biomicroscope should be
directly in front of the patient's eye, the
illumination source at about 45 degrees and the
illumination mirror in "click" position. The
slit width is almost closed, about 0.25 mm wide
and 7 to 9 mm high. First place the
magnification on low to medium (7 - 10 X) and
focused on the patient's closed lid. The
thickness of the eyelid is about 1 mm thick,
which means focusing on the cornea is
accomplished by only slightly moving the
joystick forward. Have the patient open their
eyes, give the patient a point of fixation such
as the fixation light, part of the
biomicroscope, or the top of the your opposite
ear. Once the cornea is in sharp focus, try
scanning the cornea from temporal limbus to
nasal limbus. To maintain a clear,
distortion-free view, the illumination source is
always moved to the opposite side when crossing
the mid-line of the cornea about at the center
of the pupil.
With
a clearly focused optic section slightly
temporal to the center of the cornea, increase
the magnification to 16x then to 20x and
illumination brightness and note the
following:
- 1)
The Front Surface Bright Zone Is The Surface
Of The Tears
2) The Next Darker Gray Line Is The
Epithelium Layer
3) The Next Brighter Thin Line Is Bowman's
Membrane
4) The Gray Wider Granular Area Is The Stroma
Zone
5) The Last Bright Inner Zone Is The
Endothelium
To
attain an optic section of the crystalline lens,
the angular separation of the illumination
source is reduced until the light beam just
grazes the edge of the pupil and the vertical
height can be reduced to approximate the pupil
size. This alignment can easily be accomplished
from outside the biomicroscope. When the beam
cuts just across the edge of the pupil, the
crystalline lens will appear sectioned. By
focusing the biomicroscopes with one hand and
controlling the direction or angle of the light
source with the other hand, the different layers
of the lens can be brought into focus; hence,
the anatomical location of any opacity can be
determined. Furthermore, the degree of nuclear
opalescence and color can be evaluated and
graded via the, lens opacities classification
system II (LOCS II)[Chylack, 1989].
Different magnifications may be used, but medium
and high give the best detail.
Van
Herick's technique for grading the anterior
chamber angles uses an optic section placed near
the limbus with the light source always at 60
degrees. The biomicroscope is placed directly
before the patient's eye. This technique only
allows you to judge the temporal and nasal
angles.
Van Herick Angle Estimation
Method
|
|
Angle Grades
|
Risk of Angle Closure
|
Cornea to Angle Ratio
|
|
4
|
Wide Open Angle;
Incapable of Closure. Iris to Cornea
Angular Separation Equals
35-450.
|
Anterior
Chamber Depth (Shadow) is Equal to or
Greater Than Corneal
Thickness
|
|
3
|
Moderately Open
Angle; Incapable of Closure. Iris to
Corneal Angular Separation Equals
20-350
|
Anterior
Chamber Depth (Shadow) is Between 1/4
and 1/2 the Corneal
Thickness
|
|
2
|
Moderately Narrow
Angle; Closure Possible. Iris to
Corneal Angular Separation Equals
200
|
Anterior
Chamber Depth (Shadow) is Equal to 1/4
the Corneal Thickness
|
|
1
|
Extremely Narrow
Angle; Closure Probable. Iris to
Corneal Angular Separation Equals
100
|
Anterior Chamber
Depth (Shadow) is Equal to Less Than
1/4 the Corneal Thickness
|
|
0
|
Basically Closed
Angle; Closure is Most Emanate. Iris to
Corneal Angular Separation Equals
00
|
Anterior
Chamber Depth (Shadow) is Only a Very
Narrow Slit or no Anterior Chamber
Angle
|
|
Adapted
from: Van Herick W, Shaffer RN,
Schwartz A. Estimation of width of
angle of anterior chamber. Am J
Ophthalmol 1969;68:626-9.
|
|
Optic
section using the Van Herick Technique
to grade the anterior chamber depth.
This is a grade 1 or narrow
angle.
|
|
The
"Split Limbal Technique" allows you to
make an estimation of the superior and
inferior angles. The slit lamp and
illumination system are in click
position aligned directly in front of
the patient. The beam width is that of
an optic section, which is focused on
the limbal cornea junction thus,
splitting the cornea and limbus. The
doctor then views the arc of light
through the cornea and that falling on
the iris without the aid of the slit
lamp. The angular separation seen at
the limbus corneal junction is an
estimation of the anterior chamber
angle depth in degrees.

"Split Limbal Technique"
Which Is Observed With The
Naked Eye
|
Grade
|
Angle
|
|
+ 4 TO 4
-
|
( 45 -
350
)
|
|
+ 3 TO 3
-
|
( 35 -
200
)
|
|
+ 2 TO 2
-
|
< 20 BUT
>
100)
|
|
1 -
0
|
(
100 OR
LESS )
|
|
|
|
b.)
Conical Beam: Examination of the
anterior chamber for cells or flare
must be performed before either
dilation or applanation tonometry.
Magnification
16 - 20x and illumination (high) or
what the patient will tolerate.
Dilation often results in an increase
in the number of cells and fluorescein
used in applanation tonometry causes an
increase in flare [Schlaegel,
1982]. This type of illumination is
used to detect floating aqueous cells
and flare by the, Tyndall effect, much
like seeing dust floating in the air of
a sun filled window.
The
traditional method of locating and
grading cells and flare is to reduce
the beam to a small circular pattern
with the light source 45 to 60 degrees
temporally and directed into the pupil.
Position the biomicroscope directly in
front of the patient's eye with as
bright illumination as the patient will
permit and high magnification. The
examiner always allows themselves a
period of time to dark adapt. The
conical beam is focused between the
cornea and the anterior lens surface
and observation is concentrated on the
dark zone between the out of focus
cornea and lens. This zone is normally
optically empty and appears totally
black. Flare (protein escaping from
dilated vessels) makes the normally
optic empty zone appear gray or milky
when compared to the uninvolved eye.
Cells (white blood cells escaping from
dilated vessels) will reflect the light
and be seen as white dots. The
following techniques have been
suggested: either to oscillate the
light source with the joy stick from
left to right while focused in the
anterior chamber or to focus from the
posterior cornea to the anterior lens
while oscillating the light source.
These techniques are not typically used
clinically
The
following is a more traditional
technique and one that works well
clinically and is superior when grading
the severity of inflammation. Use a
parallelepiped approximately 2 mm wide
and 4 mm high. Focus on the iris near
he pupil then pulled back the focus
into the anterior chamber so light is
seen passing through the out-of-focus
cornea and lens. The examiner waits and
watches the dark zone between the
out-of-focus cornea and the light
passing through the pupil and lens. The
convection currents of the aqueous will
move any protein or cells up and
through this zone. You watch and count
the number of cells seen during a
minute period.
|
Grading Cells and Flare
|
Grade
|
Aqueous Cells
|
Grade
|
Flare
|
|
0
|
None
|
0
|
Optically Empty Compared
Bilaterally
|
|
1
|
2-5 Cells Seen in 45 Seconds
or One Minute
|
1
|
Faint: Haze or Not Equal
Bilaterally
|
|
2
|
5-10 Cell Seen at Once
|
2
|
Moderate: But Iris Detail
Still Clear
|
|
3
|
Cells Scattered Through Out
Beam 20 or More
|
3
|
Marked: Iris Details Becoming
Hazy
|
|
4
|
Dense Cells in Beam, More
Than You Can Count
|
4
|
Dense Haze: With Obvious
Fibrin Collecting on Iris
|
READ VOL. 4 - CHAPTER 32 IN "DUANES' CLINICAL
OPHTHALMOLOGY"
|
Cells
and flare in the anterior chamber
represent a condition of great concern
and are usually diagnostic of an
inflammation. However, if cells or
flare are not seen, but an inflammation
is suspected, use the "Consensual
Pupillary Reflex" test "Henkind" test
or "Consensual Pain Reflex" test, all
one and the same, to help confirming an
inflammatory diagnosis. The patient
completely covers the eye in question
so no light can enter. They are to
report any discomfort when the slit
lamp is turned on in front of the "good
eye" and the brightness is turned up.
If he/she reports discomfort, cells and
flare may not be present, but there
most likely is a smoldering
inflammation that has not resolved or
is about to develop [Au,
1981].
Grading the Consensual Pain Reflex
|
Grade
|
Patient
Response
|
|
1 To 1+
|
Definite Pain Without Acute
Distress
|
|
2 To 2+
|
Causes Wincing or Complaint
of Pain
|
|
3 To 3+
|
Causes Withdrawal
From the Light
|
|
4 To 4+
|
Severe Allows No Light in
the Eye
|
|
|
C.)
Parallelepiped: A parallelepiped is
one of most common types of
illumination used. It is used in
combination with a number of different
types of illuminations. The
biomicroscope should be directly in
front of the patient's eye, the
illumination source at about 45 degrees
and the illumination mirror in "click,"
position. A parallelepiped is
essentially an optic section, except
the slit width is greater (2.0 - 4.0
mm) and the height may vary, providing
a more three dimensional view of the
cornea or crystalline lens. The
three-dimensional view permits
observation of distinguishable details
within the crystalline lenses "zones of
discontinuity". As with the optic
section, the angle between the
illumination source and biomicroscope
may be varied to expose more corneal
epithelium, stroma and endothelium. The
whole cornea should be scanned using a
parallelepiped. When scanning the
cornea, a clear undistorted view must
be maintained by positioning the light
source to the opposite side when
crossing the mid-line of the cornea.
Both normal and abnormal findings can
be seen when scanning the cornea with
varied levels of magnifications and
brightness. Look for any of the
following:
- Tear
debris is usually benign and
related to allergies or sinus
conditions, but may correlate with
bacterial infections.
Corneal nerves are white
thread-like structures that
bifurcate and trifurcate and are
located anywhere within the
cornea.
Blood filled vessels extend
from the limbus onto or into the
cornea, and are diagnostic of
chronic or acute insult or
inflammation.
Ghost vessels extend from the
limbus onto or into the cornea. They
are empty of blood and diagnostic of
some type of past corneal insult or
inflammation.
Corneal scars are white in
color and diagnostic of some past
corneal damage, ulcer, abrasion or
foreign body.
Corneal striae are white
usually vertical thread-like
twisting lines found in Descemet's
membrane and posterior stroma. They
are diagnostic of poor soft contact
lens fitting, diabetes or metabolic
changes as with the reduced number
of endothelial cells of the elderly.
They are the result of overall
thickening of the entire cornea and
buckling of the back
surface.
Grades of Cornea Striae
|
Grade
|
Observed
Number
|
|
0
|
None
|
|
1
|
Less Than
Five
|
|
2
|
Five To Ten
|
|
3
|
Ten To Twenty
|
|
4
|
More Than
Twenty
|
- Endothelial
pigmentation when heavy and
located vertically on the
endothelium is known as
"Krukenberg's Spindle", it may be
diagnostic of iris atrophy and
pigmentary glaucoma.
Transillumination of the iris should
be performed and any
transillumination iris defects
(TID's), holes in the iris, noted.
Scant, very fine deposits are
commonly seen and not
pathological.
3.)
Retro-Illumination: Usually uses a
parallelepiped that bounces unfocused
light off one structure while observing
the back lighting of another. The
alignment and angular separation of the
biomicroscope to the illumination
source will vary. The light source beam
is reflected off another structure like
the iris, crystalline lens or retina
while the biomicroscope is focused on a
more anterior structure.
Retroillumination or transillumination
the iris or crystalline lens uses low
to medium magnification
(7 - 10x). The slit width 1 - 2 mm wide
and 4 - 5 mm high with the
biomicroscope and light source placed
in direct alignment with each other.
They are both positioned directly in
front of the eye to be examined. Focus
the slit just off the edge of the iris
and on the front of the lens. If there
are defects or atrophy of the iris they
will be seen as a retinal "orange" glow
coming back through each defect or
hole. Patients who have numerous
endothelial pigment deposits you must
transilluminate their iris. Remember
the term transillumination iris defects
or (TID's). See color plates in Dr.
Casser's book. Furthermore,
retroillumination of the crystalline
lens is required to classify and grade
both cortical and posterior subcapsular
cataracts using LOCS II.
The
cornea is probably the most common
structure viewed in retro-illumination.
Keratic precipitates (accumulation of
white blood cells and fibrin) will
appear white in direct illumination but
dark by retro-illumination. This
technique is valuable for observation
of deposits on the corneal endothelium
and invading blood vessels. According
to some authors this is the only way by
which tiny rod-like fibrin flecks may
be seen on the back of the cornea,
warning the so-called "quiet iritis" is
still active. Retro-illumination is
regarded by most optometrist to be
second in importance only to direct
illumination.
4.)
Sclerotic Scatter: This
illumination uses a parallelepiped at
the limbus to scatter light internally
throughout the cornea. Use low 6 - 10x
magnification. In the case of central
corneal clouding (CCC) the
biomicroscope is not used. The pupil is
observed with the naked eye from an
angle directly opposite from the light
source.
5.)
Indirect-Lateral-Proximal: Place
the biomicroscope directly in front of
the patient's eye and the illumination
light source at about 45 degrees. Make
sure the illumination mirror is in
"click" position. Use a parallelepiped
beam sharply focused on a given
structure like the cornea. The light
passes through the cornea and falls out
of focus on the iris. The dark area
just lateral or proximal to the
parallelepiped is the indirect or
proximal zone of illumination. This is
the area of the cornea which one
surveys through the biomicroscope. This
type of illumination is widely used for
observation of the corneal epithelium
and tears. Most helpful in detection of
mycrocystic edema, faint corneal
infiltrates and other types of
irregularities of the epithelium and
tears. Because it utilizes direct,
indirect and retroillumination
simultaneously, one should consider it
to be as important as any other type of
illumination.
|
|
|
Keratic
precipitates on the endothelium of the
cornea as seen in direct, indirect, and
retroillumination using a
parallelepiped.
|
|
6.)
Specular Refection: Again a
parallelepiped is used. This is the
only means by which one is able to view
the endothelial cells of the cornea or
the epithelial cells on the back of
lens. The cells are seen only by one
eye and they appear in the ocular
opposite from the direction of the
illumination light source.
The basic requirements for specular
reflection are as follows:
- 1)
The angle between the illumination
source and biomicroscope is
approximately 60 degrees.
2) High magnification must be
used.
3) High illumination is needed.
4) A parallelepiped beam of light is
used.
Place
he biomicroscope directly in front of
the patient's eye and the illumination
light source at 45 - 60 degrees. Just
off the limbus, obtain a sharply
focused parallelepiped of the cornea.
Slowly advanced it across the cornea
until a dazzling reflection of the
filament is seen within the
biomicroscope. This reflection is only
seen by one eye the other eye is not
bothered. Keeping the reflected light
within the biomicroscopes field of
view, the focus is moved back toward
the endothelial cells. There will be a
point where two images of the filament
are seen, one bright, and the other
ghostlike or copper-yellow in color.
Critically focus the biomicroscope on
the latter until a mosaic of hexagonal
cells are seen. It should be noted that
even with 40x magnification the
endothelial cells do not look as large
as most texts show. They resemble the
appearance of the dimpled surface of an
orange peel or basketball. When the
slit lamp's illumination system and the
biomicroscope are at equal angles of
incidence and reflection the cornea's
endothelium is viewable. Both front and
back surfaces of the crystalline lens
can also be viewed using specular
refection.
|
|
Positioning The Patient In
The Slit Lamp
|
|
1.)
Procedure:
Inform the patient what you are going
to do and why. For Example: This is a
slit lamp biomicroscope and I'll be
examining the general health of your
eye.
2.) Head Position:
A.) Tell the patient what you
want them to do: chin in the chin rest
and forehead up against the headrest.
For professional and hygienic reasons
always place a facial tissue on the
slit lamp's chin rest. You should have
already cleaned the head and chin rest
with an alcohol swab. This is always
done between every patient. This not
only helps keep things more antiseptic,
but also makes the slit lamp smell
clean and more professional.
B.) Make sure the patient not
only looks comfortable but is
comfortable. Their forehead tight
against the headrest, chin firmly down
on the chin rest and their outer
canthus aligned with the black marker
on the slit lamp post. At this point it
is a good idea to reach around and
gently pull their head slightly forward
against the headrest.
3.) Fixation Instructions:
The patient must be given fixation
instructions, where you want them to
look. This might be the fixation light,
part of the slit lamp or just past your
ear.
4.) Pre-Alignment And
Focusing:
Tell your patient to close their eyes
and relax while you get things aligned.
Turn the biomicroscope on and focus the
light source on the patient's lid. The
eyelid is only about 1 mm thick,
therefore, when you instructed the
patient to open their eyes you should
almost be in focus on the tear film of
the cornea.
|
Suggested Slit Lamp Examination Procedure
|
(1)
|
(2)
|
|
|
|
Using a broad beam
or better a 2 to 4 mm wide
Parallelepiped Type Illumination,
Magnification 10-16x, Illumination on
low at 45 Degrees, Examine both the
upper and lower lids and lashes. The
patient's eyes are open and the
illumination source is moved at the
midline of the lid.
|
Have the patient
look to their left, light source to
your left at approximately 45 degrees.
The microscope is set straight ahead.
Scan and examine the temporal bulbar
conjunctiva
|
|
(3)
|
(4)
|
|
|
|
Have the patient
look to their right, light source to
your right at approximately 45 degrees.
The microscope is set straight ahead.
Scan and examine the nasal bulbar
conjunctiva.
|
Have the patient
look up, retract the lower lid, examine
the lower bulbar, lower Palpebral
conjunctiva and inferior cornea. The
light source should be moved across to
the opposite side at the midline of the
eye. The microscope is set straight
ahead.
|
|
(5)
|
(6)
|
|
|
|
Have the patient
look down, retract the upper lid,
examine the upper bulbar conjunctiva
and superior cornea. The light source
should be moved across to the opposite
side at the midline of the eye. The
microscope is set straight
ahead.
|
Use a
Parallelepiped, 16X magnification,
light source at 45 degrees and the
microscope set straight ahead. Scan and
examine the cornea. The light source
should be moved across at the midline
of the cornea.
|
|
(7)
|
(8)
|
|
|
|
Use a full length
Optic Section, magnification 16X, light
source at 60 degrees and the microscope
set straight ahead. Evaluate and grade
the temporal and nasal angles using the
Van Herick Technique.
|
Use a narrow
Parallelepiped, 16X Magnification,
light source at 45 degrees and the
microscope set straight ahead. Examine
the iris, crystalline lens and the
anterior vitreous body.
|
|
Important:
Always, pull the slit lamp back,
shut off the instrument, and lock it
down at the end of any
procedure.
The above is only intended as a
schematic and students are encouraged
to develop any order with which they
feel comfortable. It should be pointed
out that all steps in the schematic are
relevant and should be part of the
procedure.
|
To
View Part Two

Return
To Home Page
|