Fluress©
will be used in this procedure, it contains a stain and
anesthetic, which allows applanation of a defined area of
the cornea with a variable force. Goldmann tonometry is
the last thing you do before dilation. Why? Because the
topical anesthetic will soften the cornea and can make
vision blurry. However,this softening allows for faster
penetration and transmission across the corneal barrier
of the mydriatics, hence, faster and more complete
dilation. The applanation of the cornea, which is ever so
light, will slightly distort the cornea and for a short
period cause blurred vision. The main point being
addressed is, "make sure you know the patient's final
Rx," otherwise it will be too late to refine after
Goldmann tonometry. Goldmann tonometry is the "Gold
Standard" by which all other tonometry is compared. The
upper "suspect reading" for Goldmann tonometry is equal
to or greater than 21 mmHg.
Fluress©
|
Contains
|
Purpose
or Action
|
|
0.25%
Sodium Fluorescein
|
Staining
Fluorescent Agent
|
|
0.4%
Benoxinate HCl
|
Anesthetic
|
|
1.0%
Chlorobutanol
|
Preservative
|
|
15%
Povidone
|
Wetting
agent and stabilizer, preventing precipitation
when Sodium Fluorescein and Benoxinate HCl are
combined together in a solution
|
NOTE:There
are generic brands of Fluress© that contain
proparacaine HCl, but have thimerosal as a
preservative. Thimerosal must be avoided it causes a
number of allergic reactions (e.g., diffuse superficial
punctate keratitis, corneal infiltrates, lid and
conjunctival edema, conjunctival injection,
etc.).
AREA
OF APPLANATION
The
diameter of the circular zone of applanation is 3.06 mm.
Hence, one can calculate the area of that circular zone
of applanation.
Area
=
(r)2
= 3.1416 x (1.53)2 = 7.3542
mm2
This has
been found to be the minimal area of applanation needed
to give accurate results, yet, causing only an increase
of 2.5 percent in intraocular pressure. Example: 20 mmHg,
which is just 1 mmHg below the suspect reading of 21 mmHg
(20 mmHg x 2.5% = 0.5 mmHg). Accordingly, the intraocular
measurements should be accurate to within + or - 0.5 mmHg
for intraocular pressures of 20 mmHg or less. Remember,
tonometry findings are not always diagnostic of
glaucoma, but just one of several findings you must
consider. There is a condition called ocular
hypertension, when the patient's intraocular
pressures (IOP) run over 21 mmHg, but have no optic nerve
head or visual field changes. These patients indeed need
close monitoring for any changes in their fields or optic
nerve heads.
Visual
fields and careful evaluation of the optic nerve heads
and the retinal nerve fibers are extremely important in
diagnosing glaucoma. If I had to pick one of the above as
the most important finding in predicting the possibility
of a patient having glaucoma I would have to say the
appearance of optic nerve heads and retinal nerve fiber
layer. One cannot be too dogmatic for a patient's visual
fields, optic nerve heads, and nerve fiber layer are so
very closely related.
- ORDER
OF DIAGNOSTIC IMPORTANCE
1.) Appearance of the optic nerve heads and nerve
fiber layer
2.) Visual fields
3.) Intraocular pressures (IOP's)
It is
somewhat interesting how the area of applanation relates
to the calculation and conversion from grams to mmHg.
Pressure is equal to force per unit area but one must
factor in the density of mercury (Hg) which is 13.6 gm/cm3.
One could say 13.6 gm/cm3 of Hg.
Conversion
to mmHg
Pressure = Force/Area
Pressure = 1 gm / .073541714 cm2 x 13.6
gm/cm3 of Hg
Pressure = 1 gm x cm3 of Hg / .073541714
cm2 x 13.6 gm
Pressure = 1 cm of Hg / 1
Pressure = 1 cm of Hg = 10 mmHg
Therefore,
one gram of force is needed to flatted an area equal to
.073541714 cm2 when the eye contains a
pressure of 10 mmHg and 2 gms when 20 mmHg, etc. Each
large number on the micrometer is in grams; 1 gram equals
10 mmHg while each small mark in between is equal to 2
mmHg.
CLEANING
& STERILIZING THE PROBE
The FDA and
CDC recommendations are in your manual, however, there
are some factors to be considered such as the
following:
(1) 3% H2O2 will ruin the probes in
a short period of time and if not properly dried can
cause marked pain and a secondary anterior uveitis that
is very difficult to resolve. Minimum of 5 minutes
maximum of 10 minutes.
(2) 70% Isopropyl alcohol will also destroy the probes
and if not properly dried can also cause marked pain and
a secondary anterior uveitis that is very difficult to
resolve.
When Using
Either Of The Above Recommended Procedures Make Sure You
Always Rinse The Probe Thoroughly With Saline Solution
Then Dry It Completely Before Using It On The Patient's
Eye. Never Leave The Probe In The Hydrogen Peroxide For
Longer Than 10 Minutes.
TONOMETRY
PROCEDURE
It is
important you inform the patient what you are going to do
and why. You do not have to go into great detail other
than you are going to check the pressure of their eyes or
simply "this is the blue light glaucoma screening test".
If the patient should ask if you are going to
touch their eye reply, I am going to use drops to numb
the front surface of your eye, the instrument will gently
touch the tear film of your eye and because of the drops
you will feel nothing. Once you have completed the
procedure you need to tell the patient your findings,
e.g., "The test indicates the pressures are
normal."
- (1) Set
the measuring drum on one (1 gm) equals 10 mmHg make
sure to align the white line on the probe carrier with
the zero or 180 degree marker of the probe.
(2) Place the light source 60 to 65 degrees temporal
to the probe.
(3) Use wide open diffuse illumination, bright
illumination, cobalt blue filter and 16x
magnification.
(4) This should go without saying, but you always
align the patient just like you would before starting
a normal slit lamp examination.
(5) Make sure the patient keeps their head tightly
against the head rest. Complete steps 8 & 9 then
(pull the slit lamp back, shut it off, and lock the
slit lamp in place.)
(6.) Always ask the patient if they have ever had a
reaction to Novacaine; normally this would be asked
during the case history. If they are you should use
Proparacaine HCl and a fluorescein strip. If not place
one drop of Fluress© O.U. an give the patient a
tissue to blot their eyes. Place the bottle on a flat
surface away from you and the patient. If you spill or
drip Fluress© on yourself or the patient's
clothing it is almost impossible to get out.
(7.) Always tell the patient not to rub their eyes.
They may blot, but not rub for a least 20 minutes.
(8.) The optical system and tonometer are "not
aligned" and must be offset by 5 to 10 degrees. The
view of the probe and mires are seen only with one
eye.
(9.) Pre-alignment is done from outside the slit lamp,
tell the patient where you want them to look.
(10.) Just before you are ready to move the slit lamp
forward have the patient blink then open their eyes
very wide. When the probe just touches the cornea
you'll see a limbal glow, stop, you have advanced it
far enough. You will see the mires through the slit
lamp at this point.
(11.) If, however, you advance the probe too far
forward or have too high of magnification you may see
nothing.
(12.) If the mires are too wide the readings will not
be accurate. Pull the slit lamp back dry off the end
of the probe. (Lashes in contact with probe, too much
Fluress© or excessive tearing are the most likely
cause) then start over.
- You do
not have to pull the probe off the patients eye to
make fine adjustments for alignment; this can be done
with small movements while on the cornea. Removing the
probe from the patients eye then putting it back on
until you have proper alignment will result in drying
and staining of the cornea.

-
-

-
Recording
Your Findings
OD 18 mmHg
= right eye
TA
Goldmann @ 9:30 a.m. i gtt Fluress© O.U.
OS 18
mmHg= left eye
If a
patient wears soft lenses you should flush the eyes out
very well, cleaning the patient up to remove all the
fluorescein, then advise them not to wear their lenses
for at least one (1) hour. Fluorescein will be absorbed
into the soft lens turning it yellow. Patients who wear
hard lenses should be advised not to wear their lenses as
long as they would normally. Topical anesthetics soften
the cornea's epithelium making an abrasion more likely.
All patients should have has much of the fluorescein dye
cleaned off as possible.
If the
corneal astigmatism is greater than 3.00 Diopters the
measurement is made 43 degrees from the meridian of the
lower power. It has been found that on very toric corneas
placing the probe offset by 43 degrees gives an area of
applanation closest to 7.35 mm2.
EXAMPLE:
41.00@30 degrees and 45.00@ 120 degrees
You would
place the red line on the probe carrier @ the 30 degree
mark on the probe. The red line and the white line on the
probe carrier are separated by 43 degrees.


TOPICAL
ANESTHETICS
|
Generic / Company
|
Time Of On Set
|
Duration
|
|
0.5% Proparacaine HCl
|
13 to 15 Seconds
|
15 Minutes
|
|
(Opthaine)
|
1+ Stinging Reaction
|
1+ Punctate Keratitis
|
|
0.4% Benoxinate HCl
|
15 Seconds
|
15 Minutes
|
|
(Dorsacaine)
|
2+ Stinging Reaction
|
2+ Punctate Keratitis
|
|
0.5% Tetracaine HCl
|
15 Seconds
|
15 to 20 Minutes
|
|
(Pontocaine)
|
3+ Stinging Reaction
|
3+ Punctate Keratitis
|
Keratitis:
Corneal staining and loss of epithelial cells, toxic in
nature
DIURNAL
VARIATION
Generally,
the variation in intraocular pressure over a 24 hour
period is considered to be 3 to 5 mmHg with the highest
readings being about (6:00 a.m.). However, there is
recent evidence that known glaucoma patient's highest
pressure findings are in the afternoon. Therefore, it
would be best to monitor any questionable patients. Take
diurnal pressure measurements during the day looking for
any pressure spikes with variations greater than 5 mmHg.
Example: 14 mmHg O.U. @ 8:30 a.m. and 21 mmHg O.U. @ 3:30
p.m. is diagnostic.
Differences
in pressure readings between the two eyes of 3 mmHg or
more must be questioned, this is not normal. There are
several conditions which must be considered when this
occurs.
(1) Glaucoma
(2) Retinal Detachment
(3) Uveitis-Iritis
(4) Poor Technique (This of course it not an
acceptable answer!)
Three are known conditions resulting in, possible,
diagnostic pressure differences between the two eyes.
Initially, the IOP's are lower in the affected eye. It
should be pointed out that in the later stages of a
uveitis the pressure may possibly become higher in the
affected eye.
It is not
uncommon when the probe is in contact with the eye to see
the fluorescein patterns pulsate, this is caused by the
venous pulsation within the eye. There are differences of
opinion when these pulsating mires should be interlocked
and the pressure reading interpreted. It is my opinion,
they should be interpreted when they are at the highest
point if the patient is a possible glaucoma suspect. If
the pulsation is significant, you can adjust the
micrometer so the semicircles pulsate equally to either
side of the correct endpoint. Some doctors take two
readings one at the high end and another at the low end
then average the two readings. Recording both the high
and low readings and the average.
Venous
pulsation is a very important finding and should always
be recorded, even if it has to be induced. Lack of venous
pulsation or the ability to induce it should make one
think of either (1) the intraocular pressure is higher
than the venous pressure or (2) the intracranial pressure
is higher than the venous pressure. This will be
discussed in more detail at a later date.
CONDITIONS
OF CONCERN/CAUTIONS
1.)
Patients with recurrent corneal erosions (be very
careful)
2.) Bullous
keratopathy (depending on the severity)
3.) Corneal
abrasions that are not totally healed (be very
careful)
4.)
Edematous corneas (be very careful)
5.) Band
keratopathy (depending on the severity)
Take three
(3) readings on each eye, later one (1) will usually be
enough when you are in practice. This is mainly to force
you to become more confident and competent when working
around patient's eyes. Like most things the more often
you do something the better you are at interpreting your
findings. You need to examine the corneas to see if you
caused any disruption or left a toe mark from the probe.
If there is considerable staining you should give the
patient some artificial tears to be used "prn" (or as
needed) should they notice any irritation.
Latest
information: Intraocular pressures (IOP)
following Laser In Situ Keratomileusis (LASIK) and
Photorefractive Keratectomy (PRK) are underestimated. It
depends on the patient's Pre-Op refractive prescription
and how much tissue needs to be removed. There is about
10 microns of tissue removed per diopter of refractive
error. A patient with a refractive error of 3 diopters of
myopia will have approximately 30 microns of tissue
removed. For this patient the IOP findings would be
underestimated by about 2mm Hg and more for higher
refractive errors. This underestimation has been reported
for Goldmann applanation tonometry. It is an important
new finding and appears to be related to changes in
corneal thickness. This might help explain normal tension
glaucoma where these patients may have thinner
corneas.
One last
thing, make sure you always take the time to clean all
the Fluress© or fluorescein off from around your
patients' eyes. Nothing looks worse than to see a patient
walking down the clinic hall or out of the building with
this yellow stain all around their eyes. Remember if you
get (Fluress© or fluorescein) on you or your
patient's clothing it's almost impossible to get out
(please, be careful out there).
To
View Tonometry Slides

Return
To Lecture Notes
Return
To Home Page